Sample Location Error - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

3
Troubleshooting and Test Functions

8. Sample location error

Check the plate configuration in the user interface
Ensure the right vials and plates are used
Perform an auto-alignment
Check the connector from the needle carrier to the sample transport unit
Check the sample transport connectors on the main board/sample
transport
Check the connector on the bottom of the sample transport unit
Change the needle (G1367-87200) and the seat (G1367-87101)
Change the needle carrier assembly (G1367-60010)
Change the sample transport unit (G1367-60019)
Change the main board (G1367-69500)
1100 Series WS MWS Reference Manual
103

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1377a

Table of Contents