Assembling the Main Cover
When required
Tools required
Parts required
The cabinet kit contains all parts, but it is not assembled.
NO TE
Observe the assembly instructions carefully. The main cover cannot be disassembled
CAU TI O N
once assembled incorrectly.
1 Insert the "Agilent Technologies 1100 Series" nameplate into the recess in
2 Place the top cover on the bench.
3 Press the door hinge in place
4 Press the side panels into the slots in the top cover.
5 Press the front door into the slots in the top cover.
Light protection kit installation
1 Remove the transparent front door and install the dark one.
2 Slide the transparent window out of the side panel.
3 Slide the dark window in the side panel.
1100 Series WS MWS Reference Manual
5065-4446 Cabinet kit
5042-1381 Name plate
the top cover
Repairing the Sampler
4
131