Needle/Sample Transport Assembly - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

Needle/Sample Transport Assembly

Needle drive
Needle lock
Flush port
Needle seat
Figure 34
The needle/sample transport is a multifunctional module capable of moving
the needle into various positions (such as different wells in two different
plates, different vials, needle wash position and the needle-seat position). The
active movable axes are the X-axis, the Z-axis and the theta-axis, the
vial-/plate pusher is an additional passive axis. All axes are stepper motor
driven and encoder controlled in order to have tight feedback for the axes
position. The theta and Z axes have spring loaded belt-tensioner.
1100 Series WS MWS Reference Manual
Well-plate sampler Sampling Unit
Introduction to the Well-plate Sampler
Analytical head
Peristaltic pump
Injection valve
6