Repairing the Sampler
9 Attach the 5/16 inch wrench to hold position at the needle assembly. Use
the 4 mm wrench to loosen the fitting of the loop capillary.
Do not bend the sheet metal of the needle.
NO TE
10 Pull the loop capillary out from the needle assembly.
1100 Series WS MWS Reference Manual