Agilent Technologies G1367A Reference Manual page 147

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

Installing the MTP main board
1 Install the new board. Ensure the ribbon cables are positioned in the slot in
2 Reconnect the connectors to the board.
3 Replace the connector screws in the back of the module.
4 Replace the M4 screw at the back of the module.
5 Install the sample transport, (see
6 Install the top cover, top plate, and foam (see
7 Replace the plate tray in the tray base.
8 Turn ON the sampler, close the front door, the system will reset
If a new sampler main board is installed, update the serial number information of the
NO TE
sampler in the user interface, see procedure bellow.
9 Check the firmware revision of the sampler. If the firmware revision is
10 Perform a "Sample Transport Self Alignment".
1100 Series WS MWS Reference Manual
the board.
page 133)
page 132).
automatically.
older than the current firmware revision of the autosampler, update the
firmware, see
"Replacing the Autosampler Firmware"
Repairing the Sampler
"Installing the sample transport"
"Top Cover and Foam"
on
on
on page 155.
147
4