Vessel Or Wash Vessel Error - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

3
Troubleshooting and Test Functions

Vessel or Wash Vessel Error

The needle does not reach the target position in the vial or in the vessel of the
well-plate.
The sensor behind the vial pusher in the needle carrier assembly detects the
successful completion of the needle movement to the vessel. If the needle fails
to reach the end point, the sensor fails to recognize the needle movement and
the error message is generated.
Probable Causes
• Bad vessel definition in the plate configuration.
• Closing mats to rigid/thick.
• Bad X or Theta positioning.
• Defective encoder on the needle carrier assembly.
Suggested Actions
86
Check the vessel definition in the plate configuration.
Check that the closing mats is not too thick.
Exchange the needle carrier assembly.
Change the sample transport assembly.
Change the MTP main board
1100 Series WS MWS Reference Manual