Step 4
Symptom
The vapor blowout period does not end approximately 2 minutes after turn-on,
initialization does not begin.
Possible causes
• The front cover is not closed
• The front cover is closed, but the user interface displays a "front cover
• The left side cover is removed
• The left side cover is installed, but the user interface displays a "side cover
Suggested actions
1100 Series WS MWS Reference Manual
"Problem with the Vapor blowout period"
open" message
open" message
Close the front cover. If the "front cover open" message is still displayed:
Make sure the front cover magnet is close enough to the magnet sensor
Check the ribbon cable from the SLS board to the main board
Replace the SLS board (G1367-66505)
replace the main board (G1367-69500)
Install the side door. If the "side door open" message is still displayed:
Make sure the side cover magnet is present in the side door
Make sure the side cover magnet is oriented with the correct pole toward
the magnet sensor on the main board and positioned in the side cover
such that it is close enough to the magnet sensor
Replace the main board (G1367-69500)