Power Supply - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

4
Repairing the Sampler

Power Supply

When required
Tools required
Parts required
The MTP board is sensitive to electrostatic discharge. Always use the ESD strap (see
CAU TI O N
"Using the ESD Strap"
Removing the power supply
1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly, (see
6 Remove the sampling unit, (see
7 Remove the MTP main board, (see
8 Remove the fan, (see
9 Remove the bottom foam.
10 Remove the exhaust fan, (see
11 Remove the power switch light pipe from the coupler.
12 Remove the two screws which hold the power supply from the rear panel.
13 Lift the power supply out of the module.
14 Remove the coupler from the switch of the power supply.
152
When defective
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
0950-2528 (rev. G and higher only) Power supply
on page 107) when handling electronic boards.
Needle/Seat" function.
page 132).
page 133).
page 146).
"Removing the sample transport"
"Removing the sampling unit"
"Removing the MTP main board"
"Removing the main fan"
"Removing the exhaust fan"
1100 Series WS MWS Reference Manual
"Top Cover and Foam"
on page 134).
on page 144).
on page 145).
on
on
on