Peristaltic Pump Error
The peristaltic pump motor in the well-plate sampler has failed.
The current on the motor is used by the MTP board to monitor the speed of the
peristaltic pump motor. If the current falls below a certain value, the error
message is generated.
Probable Causes
• Defective motor.
• Defective SUD board.
• Defective MTP main board.
Suggested actions
1100 Series WS MWS Reference Manual
Exchange the peristaltic pump motor.
Exchange the SUD board.
Exchange the MTP main board.
Troubleshooting and Test Functions
3
85