Agilent Technologies G1367A Reference Manual page 267

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

8
Control Module Screens of the Well-plate Samplers
EMF (Early Maintenance Feedback)
Use the F1-key (EMF) to set EMF limits. Choose Setup limits to select the
number of injection valve cycles or number of needle movements into the
needle seat at which you want to receive a warning. Changes must be
acknowledged with F6 (Done).
If a set limit has been exceeded, a message box will pop up. If you press Reset,
the limits will be removed. Ignore will continue to keep the EMF flag set.
An exceeded limit will not stop a sequence or run (information only, to plan maintenance
NO TE
activities).
1100 Series WS MWS Reference Manual
267