Agilent Technologies G1367A Reference Manual page 275

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

Table 80
Type
GLP features
Communications
Safety features
Injection range
Precision
Sample viscosity range
Sample capacity
Injection cycle time
Carry-over
1100 Series WS MWS Reference Manual
Performance Specifications Agilent 1100 Series Micro Well Plate Sampler
Specification
Early maintenance feedback (EMF), electronic records of
maintenance and errors
Controller-area network (CAN). RS232C, APG-remote standard,
optional four external contact closures and BCD vial number output
Leak detection and safe leak handling, low voltages in maintenance
areas, error detection and display
0.01 – 8 µl in 0.01 µl increments with the small loop capillary
0.01 – 40 µl in 0.01 µl increments with the extended loop capillary
Typically < 0.5 % RSD of peak areas from 5 – 40 µl,
Typically < 1 % RSD from 1 – 5 µl
Typically < 3 % RSD from 0.2 – 1 µl
0.2 – 5 cp
2 × well-plates (MTP) + 10 × 2 ml vials
100 x 2 ml in one tray
40 x 2 ml in half tray
Typically < 30 s using following standard conditions:
Default draw speed: 4 µl/min
Default eject speed: 10 µl/min
Injection volume: 0.1 µl
Typically < 0.05 % using the following conditions:
Column: 150 x 0.5 mm Hypersil ODS, 3 µm
Mobile phase: Water/Acetonitrile = 85/15
Column Flow rate: 13 µl/min
Injection volume: 1 µl caffeine (=25ng caffeine), 1 µl water to test
carryover
Outside wash of needle before injection: 20 sec with water using
flush port
9
Specifications