Installing The Loop Capillary - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

4
Repairing the Sampler

Installing the Loop Capillary

When required
Tools required
Parts required
When opening capillary or tube fittings, solvents may leak out. Please observe
WA RN ING
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
1 Pull the new loop capillary in the needle assembly.
2 Attach the 5/16 inch wrench to hold position at the needle assembly. Use
3 Push the loop capillary into the loop capillary protection tube until it comes
4 Re install the loop capillary corrugated waste tube over the loop capillary.
5 Retighten the loop capillary fitting on the analytical head.
6 Pinch the holder clamp and reinsert the needle assembly into the needle
7 Push the black chain into the needle assy until the stop.
8 Lock the needle tighter lock system.
9 Check the alignment of the needle in the needle pusher of the needle carrier
The needle must be centered in the needle pusher as all alignment by the well-plate
NO TE
sampler is calculated from the needle pusher position.
126
Capillary blocked
Capillary broken
Two 1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
G1367-87300 Loop capillary (injection volume up to 100 µl) for the G1367A/68A
G1375-87315 Loop capillary (injection volume up to 8 µl) for the G1377A/78A
G1377-87300 Loop capillary (injection volume up to 40 µl) for the G1377A/78A
the second wrench to tighten the loop capillary fitting.
out on the sampling unit side.
carrier.
by viewing from several directions to see that it is aligned in the center of
the needle pusher.
1100 Series WS MWS Reference Manual