Installing The Injection Valve Assembly - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

4
Repairing the Sampler

Installing the injection valve assembly

1 Feed the cable of the new valve through to the SUD board.
2 Connect the valve cable to the connector labelled "valve assy".
3 Install and tighten the three screws which fix the injection valve assembly.
4 Install the sampling unit, (see
5 Install the sample transport (see
6 Install the top cover, top plate, and foam (see
7 Replace the plate tray in the tray base.
8 Turn ON the sampler, close the front door.
9 The system will reset automatically.
10 In the user interface perform a "Sample Transport Self Alignment".
140
page 133).
page 132).
"Installing the sampling unit"
"Installing the sample transport"
"Top Cover and Foam"
1100 Series WS MWS Reference Manual
on page 135).
on
on