Installing the Sampler

Flow Connections to the Sampler

Preparation
Parts required
When opening capillary or tube fittings, solvents may leak out. Please observe
WA RN ING
appropriate safety procedures (for example, goggles, safety gloves and protective
clothing) as described in the material handling and safety data sheet supplied by the
solvent vendor, especially when toxic or hazardous solvents are used.
1 Connect the pump outlet capillary to port 1 of the injection valve.
2 Connect column-compartment inlet capillary to port 6 of the injection
3 Connect the corrugated waste tube to the seat adapter and the solvent
4 Ensure that the waste tube is positioned inside the leak channel.
5 Drive the tube from the peristaltic flush pump to the solvent bottle in the
6 Seat capillary: see recommendations in
30
Sampler is installed in the LC system
Parts form the accessory kits, see
valve.
waste from the leak plane.
solvent cabinet
page 54
"Accessory Kits"
on page 18
"Choice of Seat Capillary"
1100 Series WS MWS Reference Manual