Installing the Needle Carrier Assembly

When required
Tools required
Parts required
1 Install a new needle carrier (G1367-60010) on place
2 Install the three holding hex screws with the 2 mm hex key.
3 Plug in the flex board on the sample transport.
4 Pinch the holder clamp and reinsert the needle assembly into the needle
5 Check the alignment of the needle in the needle pusher of the needle carrier
The needle must be centered in the needle pusher as all alignment by the well-plate
NO TE
sampler is calculated from the needle pusher position.
6 Remove the silicon safety tube from the needle.
7 Replace the plate tray in the tray base.
8 Re-install the side door and close the front door.
9 In the user interface close the "Change Needle Carrier" function and exit
1100 Series WS MWS Reference Manual
When the needle carrier is defect
2 mm hex key 8710-2438 (supplied in accessory kit)
G1367-60010 Needle Carrier assembly
carrier.
by viewing from several directions to see that it is aligned in the center of
the needle pusher.
the maintenance mode. The instrument will reset.
Repairing the Sampler