The Main Power Supply Assembly - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

7
Theory of Operation

The Main Power Supply Assembly

The main power supply comprises a closed assembly (no on site repair
possibility).
The power supply provides all DC voltages used in the module except for the
voltages supplied by the lamp power supply to the deuterium and tungsten
lamps in the detectors. The line voltage can vary in a range from 100 240 volts
AC ± 10 % and needs no manual setting.
Line
voltage
Figure 45
To disconnect the instrument from line, unplug the power cord. The power
supply still uses some power, even if the power switch on the front panel is
turned off.
238
MPS
primary
circuits
sense
(*) total power consumption on +36 V and +24 V must not exceed 107 watts.
Main Power Supply (MPS) Blockdiagram
+12 V
supply
+24 V
supply
+5 V
switching
linear
voltage
regulators
1100 Series WS MWS Reference Manual
+36 V / 2.5 A (*
+24 V / 4.5 A (*
+5 V / 3 A
+15 V /.3 A
-15 V /.3 A