SLS Board

When required
Tools required
Parts required
Removing the SLS board
1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly,
6 Remove the sampling unit, (see
7 Remove the SLS board cover.
8 Disconnect the ribbon cable from the SLS-board and unscrew the 2 holding
9 Remove the SLS board from the inside of the module.
Installing the SLS board
1 Install the new SLS board and the SLS board cover.
2 Install the sampling unit, (see
3 Install the sample transport, (see
4 Install the top cover, top plate, and foam (see
5 Replace the plate tray in the tray base.
6 Turn ON the sampler, close the front door, the system will reset
7 In the user interface perform a "Sample Transport Self Alignment".
1100 Series WS MWS Reference Manual
When defective.
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
G1367-66505 SLS board
Needle/Seat" function.
page 132).
page 133).
screws.
page 133)
page 132).
automatically.
"Top Cover and Foam"
(see"Removing the sample transport"
"Removing the sampling unit"
"Installing the sampling unit"
"Installing the sample transport"
"Top Cover and Foam"
Repairing the Sampler