Exhaust Fan
When required
Tools required
Parts required
Removing the exhaust fan
1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly,
6 Remove the MTP main board, (see
7 Remove the exhaust fan.
Installing the exhaust fan
1 Install the exhaust fan.
2 Install the MTP main board, (see
3 Connect the fan connector to the MTP main board.
4 Install the sample transport (see
5 Install the top cover, top plate, and foam (see
6 Replace the plate tray in the tray base.
7 Turn ON the sampler, close the front door, the system will reset
1100 Series WS MWS Reference Manual
When defective.
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
3160-4097 Fan
Needle/Seat" function.
page 132).
page 133).
page 146).
page 147).
page 133).
page 132).
automatically.
Repairing the Sampler
"Top Cover and Foam"
(see"Removing the sample transport"
"Removing the MTP main board"
"Installing the MTP main board"
"Installing the sample transport"
"Top Cover and Foam"
4
on
on
on
on
on
on
145