10 Remove the silicon safety tube from the needle.
11 Replace the plate tray in the tray base. Re-install the side door and close the
12 In the user interface close the "Change Loop Capillary" function and exit
1100 Series WS MWS Reference Manual
front door.
themaintenance mode.
Repairing the Sampler
4
127