Peristaltic Pump - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

4
Repairing the Sampler

Peristaltic Pump

When required
Tools required
Parts required
The peristaltic pump is a replaceable unit. The tubing inside the pump is not replaceable.
NO TE
1 Remove the corrugated leak tubing.
2 Press the two clips on the front of the peristaltic pump.
3 Pull the pump forward off the motor shaft.
4 Disconnect the tubing leading to the wash port and the tubing coming from
5 Connect the wash port tubing to the upper tubing of the new pump (use
6 Connect the tubing coming from the solvent bottle to the lower tubing of the
7 Push the pump onto the motor shaft until the clips click into place.
8 Reinstall the corrugated leak tubing.
128
Tubing blocked or broken.
sand paper
5065-4445 Peristaltic pump
the solvent bottle.
sandpaper to get a good grip on the tubing).
new pump.
1100 Series WS MWS Reference Manual

Advertisement

Table of Contents
loading

This manual is also suitable for:

G1377a

Table of Contents