1
Installing the Sampler

Site Requirements

A suitable site environment is important to ensure optimal performance of the
autosampler.
Power Consideration
The autosampler power supply has wide-ranging capability (see
page 16). Consequently there is no voltage selector in the rear of the
autosampler. There are also no externally accessible fuses, because automatic
electronic fuses are implemented in the power supply.
The thermostatted autosampler comprises two modules, the sampler (G1367A
or G1377A) and the thermostat (G1330A). Both modules have a separate
power supply and a power plug for the line connections. The two modules are
connected by a control cable and both are turned on by the sampler module.
The thermostat power supply has two externally accessible fuses.
To disconnect the autosampler from line power, unplug the power cord. The power
WA RN ING
supply still uses some power, even if the power switch on the front panel is turned
off.
To disconnect the thermostatted autosampler from line power, unplug the power
WA RN ING
cord from the autosampler and the ALS thermostat. The power supplies still use
some power, even if the power switch on the front panel is turned off. Please make
sure that it is always possible to access the power plug.
Shock hazard or damage of your instrumentation can result if the devices are
WA RN ING
connected to a line voltage higher than specified.
14
1100 Series WS MWS Reference Manual