Front Door Error

The front door and/or the SLS board are damaged.
Probable Causes
• The sensor on the SLS board is defective.
• The door is bent or the magnet is misplaced/broken.
Suggested Actions
72
Exchange the door.
Exchange the SLS board.
Exchange the MTP board.