Errors which may occur during the turn ON and initialization process

Step 1
Symptom
Possible causes
• Defective main board
• Power supply defective
Suggested actions
Step 2
Symptom
Early during the turn-on/initialization process a main fan or an exhaust fan
error occurs.
Possible causes
• Fan not connected to the main board
• Fan connector defective
• Fan defective
• Main board defective
1100 Series WS MWS Reference Manual
"Failure to turn ON"
No activity when power button is pushed on. Power indicator stays off.
Turn the WPS off. Disconnect the power supply from the main board, see if
power indicator becomes green when the WPS is turned-on.
If yes, change the main board (G1367-69500)
If no, change the power supply (0905-2528 Rev G or higher)
"Fan failure"
Troubleshooting and Test Functions