Leak Sensor Open - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

3
Troubleshooting and Test Functions

Leak Sensor Open

The leak sensor in the sampler has failed (open circuit).
The current through the leak sensor is dependent on temperature. A leak is
detected when solvent cools the leak sensor, causing the leak-sensor current to
change within defined limits. If the current falls outside the lower limit, the
error message is generated.
Probable Causes
• Leak sensor not connected to the MTP board.
• Defective leak sensor.
Suggested Actions
66
Ensure the leak sensor is connected correctly.
Exchange the leak sensor.
1100 Series WS MWS Reference Manual