Agilent Technologies G1367A Reference Manual page 123

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

Installing the
1 Install the new metering seal. Press it firmly into position.
metering seal
2 Reassemble the analytical head. Press the plunger assembly into the seal.
3 Put the two fixing screws in place and reinstall the analytical head to the
4 Tighten alternately the two fixing screws with the 4 mm hex key.
5 Connect the two capillaries to the analytical head (Use a 1/4 inch wrench if
6 Reinstall the corrugated leak tubing.
7 Close the front door.
8 In the user interface close the "Change Piston" function and exit the
1100 Series WS MWS Reference Manual
sampling unit.
you have a SST capillary or a 4 mm wrench if you have a fused silica
capillary).
maintenance mode.
Repairing the Sampler
4
123