Side Door Error

The side door and/or the MTP board are damaged.
Probable Causes
• The door is bent or the magnet is misplaced/broken.
• The sensor on the MTP board is defective.
Suggested Actions
1100 Series WS MWS Reference Manual
Change the side door.
Exchange the MTP board.