1
Installing the Sampler
Installing a Thermostatted Sampler
Preparation
Parts required
1 Place the thermostat on the bench.
2 Remove the front cover and route the condensation drain tube to the waste
Make sure that the condensation drain tube is always above the liquid level in the
WA RN ING
vessel. If the tube is located in liquid the condensed water cannot flow out of the
tube and the outlet is blocked. Any further condensation will then remain in the
instrument. This may damage the instruments electronics.
Condensation drain tube
Waste bottle
Figure 6
3 Install the LAN interface board in the sampler (if required), see
4 Remove the adhesive tape which covers the side and front doors.
5 Open the front door and remove the left side door.
6 Remove the transport protection foam.
7 Re-install the corrugated waste tube in the plastic port.
26
Locate bench space
Provide power connections
Unpack the sampler and the thermostat
Sampler and thermostat
Power cord, for the other cables see below
ChemStation and/or Control Module G1323B
bottle.
Condensation leak outlet
Board"
on page 129.
and,"Cable Overview"
1100 Series WS MWS Reference Manual
on page 178
"Interface