Repairing the Sampler

Leak Sensor

When required
Tools required
Parts required
Removing the leak sensor
1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly, (see
6 Remove the sampling unit, (see
7 Disconnect the leak sensor from the SUD board (labelled "leak sensor").
8 Remove the sensor.
Installing the leak sensor
1 Feed the cable of the new sensor through the base of the sampling unit to
2 Connect the connector to the connector (labelled "leak sensor").
3 Install the sampling unit, (see
4 Install the sample transport, (see
5 Install the top cover, top plate, and foam (see
6 Replace the plate tray in the tray base.
7 Turn ON the sampler, close the front door, the system will reset
8 In the user interface perform a "Sample Transport Self Alignment".
154
When defective
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
5061-3356 Leak sensor
Needle/Seat" function.
page 132).
page 133).
the SUD board.
page 133)
page 132).
automatically.
"Top Cover and Foam"
"Removing the sample transport"
"Removing the sampling unit"
"Installing the sampling unit"
"Installing the sample transport"
"Top Cover and Foam"