Agilent Technologies G1367A Reference Manual page 264

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

8
Control Module Screens of the Well-plate Samplers
System - Tests
Press F3 from the System View and choose the WPS to access the WPS Tests
screen.
The Tests screen gives you access to multiple tests and features such as the
automatic transport unit Alignment tool,maintenance procedures to Change
parts on the WPS or additional test Signals to monitor various parameters
during the WPS operation.
Use the F1-key (Align Trans) to perform an automatic alignment of the WPS
transport unit.
Only the empty tray base should be installed, when performing the alignment. The
WA RN ING
wellplates must be removed from the WPS tray base before performing the transport
unit alignment. If the wellplates are installed during the alignment procedure, the
transport unit might get jammed or become defective.
264
1100 Series WS MWS Reference Manual