Theory of Operation
Table 74
One start bit and one stop bit are always used (not selectable).
Per default, the module will turn into 19200 baud, 8 data bit with no parity.

Forced Cold Start Settings

Switches 1 and 2 do not force storage of this set of parameters in non-volatile
memory. Returning switches 1 and 2 to other positions (other than being both
up) will allow for normal operation.
Forced cold start erases all methods and data stored in the non-volatile memory.
CAU TI O N
Exceptions are diagnose and repair log books which will not be erased.
If you use the following switch settings and power the instrument up again, a
forced cold start has been completed.
Table 75
To return to normal operation, set switches back to your GPIB or RS 232
configuration settings.
236
Parity Settings
Switches
Parity
7
8
0
0
No Parity
1
0
Odd Parity
1
1
Even Parity
Forced Cold Start Settings
Mode Select
1
TEST/BOOT
1
2
3
4
1
0
0
1100 Series WS MWS Reference Manual
5
6
7
0
0
0
8
1