SUD Board

When required
Tools required
Parts required
Take care not to damage the flex board when removing the SUD board.
CAU TI O N
Removing the SUD board
1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly,
6 Remove the sampling unit, (see
7 Disconnect all connectors from the SUD board.
8 Remove the two fixing screws from the SUD board, and lift out the board.
1100 Series WS MWS Reference Manual
When defective.
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
G1313-66503 SUD board
Needle/Seat" function.
page 132).
page 133).
Repairing the Sampler
"Top Cover and Foam"
(see"Removing the sample transport"
"Removing the sampling unit"
4
on
on
on page 134).