6. Needle into seat error

If the needle into seat error appears during the initialization of the Well Plate Sampler:
NO TE
• Turn-off the system
• Perform a forced cold start (move the 8-bit configuration dip switches 1,2,8
• Turn-on the system
• Perform an auto-alignment
• Turn-off the system
• Move the 8-bit configuration dip switches 1,2,8 at the rear of the instrument
• Turn-on the system
1100 Series WS MWS Reference Manual
Upgrade the firmware revision to A.04.14 or higher and the ChemStation
revision to A.08.04 or higher
Check needle position and correct alignment in the pusher
Perform an auto-alignment
at the rear of the instrument to position 1)
to position 0
Check the connector from needle carrier to the sample transport unit
Check the sample transport connectors on the main board/sample
transport
Check the connector on the bottom of the sample transport unit
Change the needle carrier assembly (G1367-60010)
Change the sample transport unit (G1367-60019)
Change the main board (G1367-69500)