Metering-Drive Motor And Belt; Removing The Metering Drive Motor And Belt - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

Metering-Drive Motor and Belt

When required
Tools required
Parts required

Removing the metering drive motor and belt

1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly,
6 Remove the sampling unit, (see
7 Disconnect the motor connector from the SUD board (labelled "Metering
8 Remove the four screws which fix the motor. Remove the motor.
9 Remove the belt and place it over the shaft.
1100 Series WS MWS Reference Manual
When belt or motor defective
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
5062-8590 Metering-drive motor
1500-0697 Belt
Needle/Seat" function.
page 132).
page 133).
M").
Repairing the Sampler
"Top Cover and Foam"
(see"Removing the sample transport"
"Removing the sampling unit"
4
on
on
on page 134).