3
Troubleshooting and Test Functions
Suggested actions
Step 3
Symptom 1
Status indicator remains off, but the rest of the turn-on/initialization process
is successful.
Possible causes
• The heel of the light pipe is not positioned correctly above the LED on the
• The main board is defective
Suggested actions
Symptom 2
Status indicator remains off, and the rest of the turn-on/initialization process
fails as well.
Suggested action
96
Make sure the fan is connected to the main board correctly
Examine the fan connector for irregularities. Correct if possible
Replace the defective fan (main fan: 3160-1017, exhaust fan: 3160-4097)
Replace the main board (G1367-69500)
"Main board initialization fails"
main board
Correctly position the light pipe
Replace the main board (G1367-69500)
Refer to the possible causes and actions for step 5 bellow