Main Fan; Removing The Main Fan; Installing The Main Fan - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

4
Repairing the Sampler

Main Fan

When required
Tools required
Parts required
The MTP board is sensitive to electrostatic discharge. Always use the ESD strap (see
CAU TI O N
"Using the ESD Strap"

Removing the main fan

1 In the user interface start the maintenance mode and select the "Change
2 Switch off the sampler at the main power switch. Remove the power cable.
3 Remove the top cover, top plate, and foam (see
4 Remove the plate tray from the tray base.
5 Remove the transport assembly,
6 Disconnect the fan connector from the MTP main board.
7 Lift the fan, and carefully pull the fan out of the foam.

Installing the main fan

1 Install the new fan and connect the fan connector to the MTP main board.
2 Install the sample transport (see
3 Install the top cover, top plate, and foam (see
4 Replace the plate tray in the tray base.
5 Turn ON the sampler, close the front door.
6 The system will reset automatically.
144
When defective.
1/4 inch-5/16 inch wrench 8710-0510 (supplied in accessory kit)
4 mm open end wrench 8710-1534 (supplied in accessory kit)
Pozidrive screwdriver
3160-1017 Fan
on page 107) when handling electronic boards.
Needle/Seat" function.
page 132).
page 133).
page 133).
page 132).
"Top Cover and Foam"
(see"Removing the sample transport"
"Installing the sample transport"
"Top Cover and Foam"
1100 Series WS MWS Reference Manual
on
on
on
on