Motor Temperature

One of the motors of the transport assembly has drawn excessive current,
causing the motor to become too hot. The processor has switched off the
motor to prevent damage to the motor.
See figure for motor identification.
The processor monitors the current drawn by each motor and the time the
motor is drawing current. The current drawn by the motors is dependent on
the load on each motor (friction, mass of components etc.). If the current
drawn is too high, or the time the motor draws current is too long, the error
message is generated.
Probable Causes
• Mechanical obstruction.
• High friction in the transport assembly.
• Motor belt tension too high.
• Defective motor.
• Defective transport assembly flex board.
Suggested Actions
1100 Series WS MWS Reference Manual
Motor 0 temperature: X-axis motor.
Motor 1 temperature: Z-axis motor.
Motor 2 temperature: Theta (gripper rotation) motor.
Switch off the sampler at the power switch. Wait at least 10 minutes before
switching on again.
Ensure unobstructed movement of the transport assembly.
Exchange the sample transport assembly.
Exchange the MTP main board.
Troubleshooting and Test Functions