Installing The Sampler - Agilent Technologies G1367A Reference Manual

1100 series well-plate sampler & micro well-plate sampler
Table of Contents

Advertisement

1

Installing the Sampler

Installing the Sampler

Preparation
Parts required
To avoid personal injury, keep fingers away from the needle area during autosampler
WA RN ING
operation. Do not attempt to insert or remove a vial or a plate when the needle is
positioned.
1 Install the LAN interface board in the sampler (if required), see
2 Remove the adhesive tape which covers the side and front doors.
3 Open the front door and remove the left side door.
4 Remove the transport protection foam.
5 Re-install the corrugated waste tube in the plastic port.
6 Re-install the left side door (take care of the magnet at the back).
7 Place the autosampler in the stack or on the bench in all horizontal
8 Ensure the power switch at the front of the sampler is OFF.
9 Connect the power cable to the power connector at the rear of the sampler.
10 Connect the CAN cable to the other Agilent 1100 modules.
11 If a Agilent ChemStation is the controller, connect either
12 Connect the APG remote cable (optional) for non Agilent 1100 Series
13 Ensure the side panel is correctly installed.
24
Locate bench space
Provide power connections
Unpack the sampler
Sampler
Power cord, for the other cables see below and
Chemstation and/or Control Module G1323B
Board"
on page 129.
position.
the GPIB cable to the detector
the LAN connection to the LAN interface
instruments.
"Cable Overview"
1100 Series WS MWS Reference Manual
on page 178
"Interface