Instrument Status Indicator

The instrument status indicator indicates one of four possible instrument
conditions:
• When the status indicator is OFF (and power switch light is on), the
• A green status indicator indicates the instrument is performing an analysis
• A yellow status indicator indicates a not-ready condition. The instrument
• An error condition is indicated when the status indicator is red. An error
1100 Series WS MWS Reference Manual
instrument is in a prerun condition, and is ready to begin an analysis.
(run mode).
is in a not-ready state when it is waiting for a specific condition to be
reached or completed (for example, front door not closed), or while a
self-test procedure is running.
condition indicates the instrument has detected an internal problem which
affects correct operation of the instrument. Usually, an error condition
requires attention (for example, leak, defective internal components). An
error condition always interrupts the analysis.
Troubleshooting and Test Functions
3