Repairing the Column Compartment

Replacing the Leak Sensor or Leak Base

When required
Tools required
Parts required
Preparations for this procedure:
Switch off the column compartment, and
disconnect the power cable.
Remove column compartment from stack
and place it on the working bench.
Remove the front cover, top cover and top
foam section, see
and Foam"
on page 65.
90
If leak sensor is defective or leak base is damaged
Screwdriver Pozidriv 1 PT3
Screwdriver flat blade
Wrench 1/4 inch
Hexagonal wrench 3 mm
Leak sensor assembly 5061-3356
Leak base G1316-43101 (part of leak panel kit G1316-68700)
"Removing the Top Cover
1 Disconnect the leak sensor assembly from the processor
board and pull the cable out of the foam channel.
1100 Series TCC Reference Manual