7
Error Information
General Error Messages

Open Cover

Error ID: 0205
The top foam has been removed.
The sensor on the main board detects when the top foam is in place. If the
foam is removed, the fan is switched off, and the error message is generated.
Probable cause
1
2
3
4
78
The top foam was removed during operation. Please contact your Agilent service
Foam not activating the sensor.
Dirty or defective sensor.
Rear of the module is exposed to strong
direct sunlight.
Suggested actions
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Ensure that the rear of module is not directly
exposed to strong sunlight.
1260 Infinity QuatPump VL User Manual