Setting Up Automaticqc Ranges; Setting Up High G/L Qc Options - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

8-6

Setting up AutomaticQC Ranges

Use the AutomaticQC ranges screen to view lot information and target ranges for controls
used for AutomaticQC analysis.
When you replace the AutomaticQC cartridge, the system replaces the current target
ranges with its own target ranges. If you want to retain the ranges that you defined,
perform the following steps. You can also use this procedure to change target ranges back
to the recommended default ranges.
1. At the Setup screen, select
If your system is connected to a data management system or an LIS and you select an
option that changes target ranges, be sure to change target ranges at the data
management system or LIS.
2. Keep or change the target ranges.
If you want to...
keep the target ranges you defined
for AutomaticQC,
change the target ranges back to the
recommended default ranges,
3. Select

Setting up High G/L QC Options

Use this screen to configure demographic data entry fields for High G/L samples. During
analysis, the system can require the operator to enter selected demographic information
for High G/L samples.
NOTE:
display the High G/L option.
1. At the Setup screen, select
2. Select either or both of the demographics you want entered when performing
High G/L samples.
3. Select the Required button to make the selected demographic required.
When you select Required next to either or both of the test demographic buttons, the
operator must enter the corresponding demographic before leaving the screen.
4. Select
02087462 Rev. V
.
Save
If both Glucose and Lactate are turned off in Setup, the Analysis screen does not
If you want the operator to enter their identification when performing a High G/L
test, select
Operator ID
If you want the operator to enter the lot number of the control when performing a
High G/L Test, select
Lot Number
.
Save
Rapidlab 1200 Operator's Guide: System Configuration
QC > AutomaticQC Ranges
then...
select
Retain Ranges
a.
Deselect
Retain Ranges
off.
b.
Select
Reset Defaults
.
QC > High G/L
.
.
.
.
to turn the option
.