Flushing The Tubing For The Measurement Module Waste - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

5-56

Flushing the Tubing for the Measurement Module Waste

1. Disconnect the tubing on the reagent manifold coming from the pressure sensor
bubbler.
2. Connect the syringe tubing to the connector on the reagent manifold.
1
2
3
Figure 5-30
3. Flush the tubing with a minimum of 3 mL of water.
4. Flush the tubing with a minimum of 12 mL air.
5. Collect the water in an absorbent swab or equivalent.
6. Reconnect the tubing to the reagent manifold.
02087462 Rev. V
Connector
Pressure sensor bubbler
Water collection area
Reagent Manifold Connector
Rapidlab 1200 Operator's Guide: Maintenance