6-34
Copying Diagnostic Data
Refer to Copying Diagnostic Data‚ page 7-3.
Using the Diagnostics Screen
The Diagnostic screen has options to perform the following functions:
•
print diagnostic reports
•
perform the Waste Detector Calibration
•
perform the Screen Calibration
•
access other diagnostic screens
Performing Tests and Printing Diagnostic Reports
Use this procedure to print diagnostic reports:
1. Select
2. Enter your password, if necessary.
3. Print the appropriate report:
To print the...
Temperature report,
Last Cal report,
Measurement report,
Voltage and
Temperatures report,
Printer report,
Wavelength Cal report
(Rapidlab 1245 and
1265 systems),
02087462 Rev. V
.
Status > Diagnostics
Select...
Temperature
Prints the temperature for various system components and
the current pAtm providing the system ID, the date and time
of the report, and temperature information.
Sensors > Last Cal
Prints data for each sensor from the last 2-point or full
calibration. Refer to Using the Sensors Screen‚ page 6-44.
Sensors > Measurement
Prints an index and one snapshot of measurement readings
for all sensors.
Voltage and Temperatures
Prints the voltage and temperatures.
.
Printer
Prints all printable characters providing the system ID, the
date and time of the report, and the printer data.
CO-ox > Wavelength Cal
Rapidlab 1200 Operator's Guide: Troubleshooting
.
.
.
.
.