Further programming
Setting the real time clock
Prerequisite
The correct date and time must be set in the service PC. If not set date at C:\> by typing date
<ENTER> and time at C:\> by typing time <ENTER>.
Setting the time in the Mammomat
Main menu:
1.
2.
3.
Setting the tube-current reduction
In Mammomat, the tube-current reduction is factory-set to "ON", which means that the
tube current is reduced to approx. 50 mA during the first 100 ms of the exposure. In this
way, possible grid lines are avoided.
If the tube-current reduction is set to "OFF", grid lines may occur at short exposure times
(< 100 ms).
NOTE
Main menu:
1.
2.
3.
4.
5.
Siemens-Elema AB
Solna, Sweden
Select Configuration <↑> <↓> < ENTER>.
Select Clock <↑> <↓> < ENTER>.
Press <F2>.
With <F2> the date and time set in the service PC are taken over by the
Mammomat.
The tube-current reduction is effective only in Iontomat mode.
Select Configuration <↑> <↓> < ENTER>.
Select Miscellaneous <↑> <↓> < ENTER>.
Select Tube mA red. <↑> <↓> < ENTER>.
Tube-current reduction (TCR) at short exposure times.
TCR switch is ON/OFF.
Exposure time with 50 mA tube current is 100 ms.
Switch ON or OFF with <↓>.
Save the programmed values with <F2>.
Register 3
SP B7-230.033.07
Rev. 03
11.99
12
Page 1 of 2
Mammomat 1000/3000
SPS-UD
Installation and Start-Up
12 - 1