Rapidlab 1200 Operator's Guide: Troubleshooting

D23 Reagent Flow Error: 4, 12

Problem
Slope Cal reagent
was not detected
during calibration.
4 FD1
12 FD2

D23 Reagent Flow Error: 5, 13

Problem
RCx reagent was
not detected during
calibration.
5 FD1
13 FD2
Corrective Action
1.
Perform the highlighted test(s) in
.
Functions
2.
Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3.
If the problem persists, replace the reagent cartridge.
Perform the highlighted test(s) in
.
Functions
4.
If the problem persists, call for technical assistance.
Corrective Action
1.
Perform the highlighted test(s) in
.
Functions
2.
Perform a leak test.
Refer to Performing the Leak Test‚ page 6-37. If the test
fails, check the tubing for leaks, and that the tubing is
correctly installed in the pinch valve, the O-rings are in
place, and the spring-loaded latch is closed. For Rapidlab
1245 and 1265 systems, verify that the CO-ox sample
chamber is installed correctly.
3.
If the problem persists, replace the wash cartridge.
Perform the highlighted test(s) in
.
Functions
4.
If the problem persists, call for technical assistance.
6-61
Diagnostics > Fluidic