Rapidlab 1200 Operator's Guide: Troubleshooting

Performing the Waste Detector Calibration

Use this procedure to calibrate the waste detector for the waste bottle.
1. Select
2. Enter your password, if necessary.
3. Select
4. Remove the waste bottle, empty it, and set it aside.
5. Select
6. Fill the waste bottle with tap water to the Max line.
7. Reinstall the bottle, and wait while the system calibrates with the full bottle.
8. Remove the waste bottle, empty it, and reinstall the empty bottle.
The system returns to the Diagnostics screen when finished.
9. To return to the Status screen, select
The systems perform a wash.

Calibrating the Touchscreen

Use this procedure to calibrate the touchscreen when the screen responds
inconsistently to selects. For example, if you must select to the left or the right
of an item to select it, the touchscreen needs calibration.
1. Select
2. Enter your password, if necessary.
3. Select
The screen displays a button.
NOTE:
screen. The system automatically returns to the Diagnostics menu after 30
seconds.
CAUTION:
more than once each time it appears. Touching any other part of the screen or
touching the button more than once causes the screen to become inoperable. If this
occurs, call for technical assistance.
4. Select the button.
The button changes position on the screen.
5. Continue selecting the button again until the system returns to the Diagnostics
screen.
6. To return to the Status screen, select
.
Status > Diagnostics
.
Waste Detector
and wait while the system calibrates without the bottle.
Continue
.
Status > Diagnostics
.
Screen Calibration
If you do not want to calibrate the touchscreen, do not touch the
Do not select any other part of the screen and do not select the button
.
Return
.
Return
6-35
02087462 Rev. V