Rapidlab 1200 Operator's Guide: Troubleshooting
b. Push the clot-removal line through the fluid connector on the reagent
1
2
Figure 6-17
c. Re-connect the CO-ox pump tubing.
5. Reinstall the cartridges and close both the reagent and wash doors.
6. Perform the RCx reagent flow test:
a. Select
b. Look at the sample path for obstructions and flow problems.
7. Return to the Status screen to view the events log.
8. Verify sensor performance:
a. Perform a 2-point calibration.
b. Analyze a minimum of 2 levels of quality control material.

Using Diagnostics

NOTE:
tests.
NOTE:
screen or subscreens are displayed. When you exit from the Diagnostics screen,
the system may perform an extended calibration before you can use the system.
manifold.
Pump tubing at the fitting.
Fluid connector.
CO-ox Sample Path to Waste
Back > Fluidic Functions > RCx
Operators with level 1 or level 2 security access can access the diagnostic
The system cannot perform routine fluidic activities while the Diagnostics
.
02087462 Rev. V
6-33