D3 Codes; D3 Slope Error: Po 2 Pco; D3 Slope Error: Ph Na K Ca Cl¯ Ref - Siemens Rapidlab 1200 Operator's Manual

Table of Contents

Advertisement

Rapidlab 1200 Operator's Guide: Troubleshooting

D3 Codes

D3 Slope Error: pO
Problem
Sensor slope is
beyond predefined
limits during
calibration
operations.
D3 Slope Error: pH Na
Problem
Sensor slope is
beyond predefined
limits during
calibration
operations.
pCO
2
2
Solution
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
Check that the sensors are installed in the correct location
and are aligned, the O-rings are in place, and the
spring-loaded latch is closed.
If necessary, remove and install the sensors correctly.
If the problem persists, replace the sensor.
If the problem persists, call for technical assistance.
+
+
++
K
Ca
Cl¯ Ref
Corrective Action
Correct any reagent problems such as D23 codes before you
attempt to correct this D code.
Perform a 2-point calibration.
Refer to Performing Manual Calibrations‚ page 3-4.
Deproteinize the sample path and condition the sensors.
Refer to Deproteinizing the Sample Path‚ page 5-4.
Check the solution level in the affected sensor and look for
bubbles.
+
++
The K
, Ca
, and Cl¯ sensors should be nearly full. The pH
+
and Na
sensors should be full. If the solution level is low,
refill the sensor.
If more than one sensor or the reference sensor has the slope
error, look for salt in the vent hole and bubbles in the sensor.
Clean and remove bubbles from the reference sensor as
needed.
Check that the sensors are installed in the correct location
and are aligned, the O-rings are in place, and the
spring-loaded latch is closed.
If necessary, remove and install the sensors correctly.
Check the sensor contacts and measurement block for salt
buildup.
Check behind the sensor for fluid leaks that can come from
the reference sensor.
If salt or leaks are present, clean the sensor with water, dry,
reinstall, and perform a 2-point calibration. Refer to
Performing Manual Calibrations‚ page 3-4.
6-51
02087462 Rev. V

Advertisement

Table of Contents
loading

Table of Contents