Appendix C: System Fluids

Recommended Fill Volumes

You can analyze samples with the Rapidlab 1200 systems using the sample
collection devices listed in the following table. Always use the recommended
minimum fill volumes for the sample devices. This ensures that the systems can
aspirate sufficient sample for analysis.
Sample Device
1-mL syringe
3-mL syringe
5-mL syringe
Capillary tube
Sampling Mode
All parameters (default)
Microsample
pH and pH, Glu, Lac
tHb
All parameters (default)
Microsample
pH and pH, Glu, Lac
tHb
All parameters (default)
Microsample
pH and pH, Glu, Lac
tHb
All parameters (default)
Microsample
pH and pH, Glu, Lac
tHb
Minimum Fill Volume
100 µL (1240 systems)
150 µL (1245, 1260 systems)
200 µL (1265 systems)
200 µL for all systems in all
modes
500 µL for all systems in all
modes
90 µL (1240 systems)
140 µL (1245 systems)
125 µL (1260 systems)
175 µL (1265 systems)
55 µL (1240, 1245 systems)
95 µL (1260, 1265 systems)
35 µL (1240, 1245 systems)
95 µL (1260, 1265 systems)
100 µL (1245, 1265 systems)
02087462 Rev. V