Performing the CO-ox Pump Test

NOTE:
Use this procedure to test the CO-ox pump. The pump test moves the pump at 4 speeds
numbered 1 through 4. When you select the
number of the current speed on the screen. The test is complete when speed 4 is displayed
on the screen.
1. Select
2. Enter your password, if necessary.
3. Select
The CO-ox pump rotates at the first speed.
4. Look for obstructions and listen for any unusual noises that may indicate a problem
with the pump.
5. Select
6. Continue selecting
tested all 4 speeds.
7. Select
8. Select
9. To return to the Status screen, select
The system performs a wash.

Using the Cartridges Screen

The Diagnostics > Cartridges screen displays buttons for performing the following
actions:
ejecting the wash cartridge
ejecting the reagent cartridge
ejecting both the wash and reagent cartridges
ejecting the AQC Cartridge
02087462 Rev. V
This test only applies to the Rapidlab 1245 and 1265 systems.
Status > Diagnostics
Pumps and Valves > CO-ox Pump > Start
to advance the pump to the next speed.
Change
Change
and wait while the system returns the pump to the correct position.
Stop
.
Return
Rapidlab 1200 Operator's Guide: Troubleshooting
button, the system displays the
Change
.
.
and listening and looking for problems until you have
.
Return